Magtrol Model 5300 Three Phase Power Analyzer
Chapter 1 - Introduction
INSTALLATION
O
RIENTATION
The 5300 PA must be mounted within ± 20 degrees
from horizontal. This ensures proper operation of the
mercury shunt bypass relays.
ELECTRICAL LOAD
Use wire rated for the maximum load current and voltage
expected.
Hint: Use wire gauge large enough to ensure good
connector set screw compression on the wire leads.
Soldering may be used instead of, or in addition to the
set screw.
IEEE-488 (GPIB)
Use only high quality shielded cable conforming to the
bus standards.
INITIAL CHECKOUT
1. Make sure the circuit is completely de-energized
by removing all voltage sources.
2. Plug the 5300 PA into the 50/60 Hz power mains.
3. Switch the POWER rocker switch (red) to ON and
observe that the digit readouts flash on-off-on and
then indicate zero or a small number.
4. The VOLTS and AMPERES range switch indicator
lights will all illuminate. The voltage ranges will
sequence down from 600 through 150, and the
current ranges will sequence down from 100 through
5.
5. The 150 Volt and 5 Amp and AUTO indicators
remain illuminated. The MODE indicators will not
light during this power-on sequencing.
6. Your 5300 PA has passed the initial check.
TRANSIENT OVERLOADS
Connect an appropriate transient suppressor in parallel
with all inductive loads. Consult the suppressor vendor's
application literature for proper selection and sizing.
Damage to the 5300 can result from
excessive voltage transients generated
by unsuppressed inductive loads. This
damage is not within the scope of the
normal instrument service and is not
covered by the Magtrol Warranty.
CURRENT OVERLOAD
There are no fuses in the 5300 PA measuring circuits.
Therefore, excessive current passed through the AMPS
terminals will cause excessive internal heating and
possible unit damage.
This overload abuse is not covered by
the Magtrol Warranty.
Know your load conditions and double check all
connections. If an overload should occur, immediately
remove all power, and locate and correct the problem
before re-energizing your circuit. If a circuit breaker is
installed, it must be installed on the LOAD side of the
5300 (downstream). This will keep the low impedance
of the input line connected to the 5300 PA for surge
suppression. If the line side must also contain a breaker,
it should be delayed in operation to open after the load
side breaker has opened.