Magtrol Model 5300 Three Phase Power Analyzer
Chapter 9 - Calibration
NOTE: Calibration on the 25 and 100
Amps ranges may have to be
done at less than full range scale,
depending on your calibration
equipment.
VOLTS BALANCE
1. Go to the 5300 PA 150 VOLTS range.
2. Set the voltage calibrator for + 150 volts DC output.
Note the VOLTS display reading.
3. Reverse the calibrator polarity to -150 volts DC. If
the plus to minus display readings differ by more
than ± 2 LSD, connect a digital voltmeter at
amplifier 78B147-U4 pin 14 and adjust trim pot
78B147-R10 for zero ± 0.1 millivolt dc.
CURRENT BALANCE
1. Go to the 5300 PA 5 amp. range
2. Set the current calibrator for + 5 Amps DC output.
3. Note the AMPERES display reading. Reverse the
calibrator polarity to - 5 Amps DC. If the plus and
minus display readings differ by more than ± 2 LSD,
connect digital voltmeter at amplifier 78B147-U1
pin 14 and adjust trim-pot 78B147-R2 for zero ±
0.1 millivolt DC.
WATTS SCALE FACTOR
1. Set the volts and current calibrators for 150 VDC
and 5 Amp DC output.
2. Adjust trim pot 78B147-R18 for 750.00 WATTS
display reading.
3. Confirm AC measurement accuracy by setting the
calibrators for 150 VRMS and 5 Amps RMS at 60
Hz. The WATTS display should read close to
750.00 watts and be within the specification
accuracy. See Section 2 - Specifications.
NOTE: Adjust current calibrator's phase
shift control for zero phase angle
between volts and amps (peak the
watts indication) - see
"Calibration Verification--AC,"
Step 3.