4Magtrol Model 5300 Three Phase Power AnalyzerChapter 2 - SpecificationsDIGITAL• Processing:Integration period = 0.10 second.Display update time = 2 r
53 - Connecting the 53003 - Connecting the 53003 - Connecting the 53003 - Connecting the 53003 - Connecting the 5300Always install a properly rated ci
6Magtrol Model 5300 Three Phase Power AnalyzerChapter 3 - Connecting the 5300SINGLE PHASE AC OR DCCONNECTIONSSee figure 2.NOTE: Connections are shown
7Magtrol Model 5300 Three Phase Power Analyzer Chapter 3 - Connecting the 53003-PHASE 3-WIRE DELTA CONNECTIONSThe power dissipated in each phase (PØ)
8Magtrol Model 5300 Three Phase Power AnalyzerChapter 3 - Connecting the 53003-PHASE, 4-WIRE WYE CONNECTIONSFigure 5. 3-Phase, 4-Wire WYE ConnectionsI
94 - Operation4 - Operation4 - Operation4 - Operation4 - OperationRead Section 1 and make sure unit is connected properly(see Section 3) before procee
10Magtrol Model 5300 Three Phase Power AnalyzerChapter 4 - OperationMODE SELECTIONSFigure 7. Mode SelectionsHOLD AVG PF PH.V.MODE HOLDThe HOLD push-bu
11Magtrol Model 5300 Three Phase Power AnalyzerChapter 4 - OperationDISPLAY SELECTIONSFigure 8. Display SelectionsØ1 Ø2 Ø3 KWΣDISPLAY Ø1, Ø2, Ø3Indivi
12Magtrol Model 5300 Three Phase Power AnalyzerChapter 4 - OperationAMPERES RANGEFigure 10. Current Range Selections5 25 100 AUTO5, 25 and 100 Amperes
135 - CT/PT Installation5 - CT/PT Installation5 - CT/PT Installation5 - CT/PT Installation5 - CT/PT InstallationExternal current and potential transfo
While every precaution has been exercised in the compilation ofthis document, Magtrol, Inc. assumes no responsibility for errorsor omissions. Addition
14Magtrol Model 5300 Three Phase Power AnalyzerChapter 5 - CT/PT InstallationCHANGING OR REMOVING CT/PTVALUESTo modify a value, simply overwrite the e
156 - GPIB Communication6 - GPIB Communication6 - GPIB Communication6 - GPIB Communication6 - GPIB CommunicationGPIB (IEEE-488)The IEEE-488 or GPIB (G
167 - 5300 P7 - 5300 P7 - 5300 P7 - 5300 P7 - 5300 PA Instruction SetA Instruction SetA Instruction SetA Instruction SetA Instruction SetPC TO 5300 PA
17Magtrol Model 5300 Three Phase Power AnalyzerChapter 7 - Instruction SetThis string transmits volts, watts and amps. To get linevoltage, power facto
18Magtrol Model 5300 Three Phase Power AnalyzerChapter 7- Instruction Set'Read 5300 PA full data - Magtrol Inc., 91 rt'Primary address set t
198 - Operating P8 - Operating P8 - Operating P8 - Operating P8 - Operating PrinciplesrinciplesrinciplesrinciplesrinciplesThe Magtrol 5300 PA uses the
20Magtrol Model 5300 Three Phase Power AnalyzerChapter 8 - Operating PrinciplesVrms= dtT0t1V2Vo=10XYCosQVrms=dtt1T0V2Vrms=tdtT10V2V/F CONV.KhzVOLT.070
219 - Calibration9 - Calibration9 - Calibration9 - Calibration9 - CalibrationGENERALComplete calibration of the 5300 PA is beyond thecapabilities of m
22Magtrol Model 5300 Three Phase Power AnalyzerChapter 9 - CalibrationSince the current calibrator output is usually a lowimpedance with low complianc
23Magtrol Model 5300 Three Phase Power AnalyzerChapter 9 - CalibrationAMPERES CALIBRATION CHECKDC:1. On the 5300, press the 5 AMPERES range push-butto
iiiTTTTTable of Contentsable of Contentsable of Contentsable of Contentsable of ContentsSALES AND TECHNICAL ASSISTANCE ...
24Magtrol Model 5300 Three Phase Power AnalyzerChapter 9 - CalibrationVOLTS ZERO1. Disable the auto zero function, refer to SpecialFunctions in Sectio
25Magtrol Model 5300 Three Phase Power AnalyzerChapter 9 - Calibration NOTE: Calibration on the 25 and 100Amps ranges may have to bedone at less than
26Appendix AAppendix AAppendix AAppendix AAppendix A: Schematic Drawings: Schematic Drawings: Schematic Drawings: Schematic Drawings: Schematic D
27Magtrol Model 5300 Three Phase Power Analyzer Appendix A: Schematic DrawingsCIRCUIT BOARD 78B145 - DIGITAL READOUT & MPUU64Q3Q69274HC3731Q5874D
28Magtrol Model 5300 Three Phase Power AnalyzerAppendix A: Schematic DrawingsCIRCUIT BOARD 78B146 - ANALOG INPUT BOARDU9x200x100914x1010151612113-78-
29Magtrol Model 5300 Three Phase Power Analyzer Appendix A: Schematic DrawingsCIRCUIT BOARD 78B147 - CONVERTER BOARDR2420K+15VC13.001U5Vo+V12+IN14132
30Magtrol Model 5300 Three Phase Power AnalyzerAppendix A: Schematic DrawingsCIRCUIT BOARD 78B149 - SWITCHBOARDC1R3200R4200R1200R2200J4SW3FRONT VIEWS
31Magtrol Model 5300 Three Phase Power Analyzer Appendix A: Schematic DrawingsCIRCUIT BOARD 78B153 - SWITCH INTERFACECD4518EN10Q314RSTCD4518EN26RSTQ3
32Magtrol Model 5300 Three Phase Power AnalyzerAppendix A: Schematic DrawingsCIRCUIT BOARD 78B154 - COUNTERS BOARDU31Q3Q2643527EN Q1RST14520Q0CKQ314Q
33Magtrol Model 5300 Three Phase Power Analyzer Appendix A: Schematic DrawingsCIRCUIT BOARD 78B175BR1PD05T114A-20-109467121013C4.1uFC31uFC14700uFC21u
ivFigure 4. 3-phase, 3-Wire WYE Connections ... 73-Phase,
Magtrol, Inc. warrants its products to be free from defects in material and workmanship under normal use andservice for a period of one (1) year from
v"SGL" INSTRUCTION ...
viThis page intentionally left blank.
1ABBREVIATIONSPA - Model 5300 Power Analyzer.GPIB - IEEE-488 Instrument Bus Standard.LED - front panel indicator light.A, V, W, PF and VA: Amperes, Vo
2Magtrol Model 5300 Three Phase Power AnalyzerChapter 1 - IntroductionINSTALLATIONORIENTATIONThe 5300 PA must be mounted within ± 20 degreesfrom horiz
32 - Specifications2 - Specifications2 - Specifications2 - Specifications2 - SpecificationsVOLTAGEThree DC and AC Ranges:• 150, 300, and 600 Volts DC