Magtrol Model 5300 Three Phase Power Analyzer
Chapter 4 - Operation
AMPERES RANGE
Figure 10. Current Range Selections
5 25 100 AUTO
5, 25 and 100 Amperes rms plus AUTO range: At power
turn on, the 5 Amps range and AUTO (auto ranging)
are active, by default. Up ranging occurs when the
measured current on any phase exceeds that range
maximum current by 5% (>105%). This feature is active
in either AUTO (LED on) or manual (LED off).
Automatic down ranging occurs only in AUTO when
the measured current falls to about ¼% below that
range's maximum value.
The PA up-ranging feature protects the instrument from
overload and ensures accurate measurements.
Exceeding the PA maximum current of 100 Amps on
any phase causes the instrument to display "HELP."
This means that you must reduce your input current
until the normal display reappears.
ANALOG OUTPUT OPTION
This option provides analog output signals for AMPS,
VOLTS and WATTS.
These three analog outputs with their common are
isolated from the measured circuits at 750 Volts
continuous breakdown.
The outputs are provided through a standard 5 pin DIN
style connector located on the rear panel of the 5300 -
see figure 11 for connections. The analog output voltage
is 5.00 volts at each AMPS, VOLTS or WATTS range
maximum value.
See Section 2 - Specifications.
NOTE: Analog output of POWER
FACTOR is not provided
Figure 11. Analog Output Connections
N/C 5 1 COM
WATTS 4 2 AMPS
3
VOLTS