Deni 5300 Specifications Page 27

  • Download
  • Add to my manuals
  • Print
  • Page
    / 40
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 26
21
9 - Calibration9 - Calibration
9 - Calibration9 - Calibration
9 - Calibration
GENERAL
Complete calibration of the 5300 PA is beyond the
capabilities of most users. Magtrol, Inc. recommends
that the 5300 PA be returned to the factory for calibration
certifying the instrument to full specifications.
However, this section includes Calibration Accuracy
Verification procedures and steps to make minor trim-
ups of calibration.
NOTE: To perform these steps, the
factory calibration seal will be
broken. This voids the factory
Calibration Certification that
comes with the 5300 PA.
TEST SETUP
The recommended test setup for each phase is shown in
Figure 16.
NOTE: The same test setup is repeated
for each of the 3-phases.
The accuracy of the voltage and current standards used
for calibration should be at least four times the accuracy
of the 5300 PA specifications. The equipment shown
in this test setup meets this requirement.
To accurately measure AC power it is important that
the current source have a phase shifting control. With
such a phase adjustment, phase shift caused by incidental
circuit inductances can be compensated for by adjusting
the phase of the current wave form to be exactly in phase
(zero phase shift) with the voltage wave form. Sources
of incidental circuit inductance include the current shunts
within the 5300, output impedance of the calibrators,
impedance of the load and wiring inductance.
OPTIONAL-FOR AMPS >2
TRANSCONDUCTANCE AMP.
FLUKE 5725A;5205A
GUIDELINE 7620A, OR
EQUIVALENT
GND
INPUT
AMPS
5300
OUTPUT
LOW
VOLTS
HIGH
EARTH
VOLTAGE CALIBRATOR
FLUKE 5700A OR
FLUKE 5100A OR
EQUIVALENT
VOLTAGE CALIBRATOR
FLUKE 5700A OR
EQUIVALENT
VOLTS VOLTS, HIGH LOW
VARIABLE
PHAS
E
OUTPUT
PHASE
LOCK
IN
LOW CURRENT
HIGH
UUT
(REF)
(EACH PHASE)
Figure 16. Calibration Verification Test Setup
(For Each Phase)
Page view 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32 ... 39 40

Comments to this Manuals

No comments