Magtrol Model 5300 Three Phase Power AnalyzerChapter 8 - Operating Principles
Vrms= dt
T
0
t
1
V
2
Vo=
10
XYCosQ
Vrms=
dt
t
1
T
0
V
2
Vrms=
t
dt
T
1
0
V
2
V/F CONV.
Khz
VOLT
.0707
U8
V/F CONV.
VOLT
.0707
Khz
.0707
V/F CONV.
Khz
VOLT
ISO
OPTO
VFo
ISO
OPTO
WFo
ISO
OPTO
AFo
I AMP
5
I AMP
25
I AMP
100
AMP
V
5A
25A
100A
V
OUTPUTS
TO
MPU
ISOLATED CIRCUIT
AMPS
INPUT
(FROM
SHUNTS)
VOLTS INPUT
(FROM
VOLTAGE
DIVIDER)
78B146-U7
78B146-U8
78B146-U9
78B146-U10
78B147-U1
78B147-U2
78B147-U7 78B147-U9 78B147-U10
78B147-U4
78B147-U5
78B147-U3
78B147-U11
78B147-U6
RANGE
Y
X
(t)
(t)
(t)
Figure 15. 5300 Block Diagram
Also, the AMPS and VOLTS signals are input to a
precision four quadrant X-Y multiplier 78B147-U7,
averaged by an active two pole Butterworth filter
78B147-U8, converted to frequency in voltage to
frequency converter 78B147-U10 and isolated by optical
coupler 78B147-U11. The resulting frequency signal,
WF
0
represents true power (V × I × Cos θ ), at a scale
factor of. 0.0707 volts per 1.0 kilohertz. There is 2500
volts isolation between the measured circuit and the AF
0
,
VF
0
and WF
0
frequency signals.
DIGITAL PROCESSING
Frequency proportional signals of AMPS, VOLTS and
WATTS (AF
0
, VF
0
, and WF
0
) are integrated for a precise
period of 0.1 seconds for digital conversion and
processing via an eight bit microcomputer (MPU). The
MPU controls functions of range control, auto-zero, fine
calibration, BCD conversion and GPIB communications.
The MPU acquisition and conversion rate is 100 milliseconds
(10 readings per second) and the display updates at 500
milliseconds (2 updates per second). The MPU scans all
front panel push-button control inputs 25 times per second.
When a push-button is pressed, the appropriate program
loop is entered and the function is executed on the next
program cycle. No damage will result to the 5300 PA from
any combination of push-button selections.
In the AVG MODE (averaging), the 5300 PA integrates
the 10 data readings per second (AMPS, VOLTS and
WATTS) and then samples the averaged value for
display. As the sampled readings accumulate, the
changing display digits will steady down to provide a
stable measurement of true power (computed from
average VOLTS and AMPS). This technique works well
when the displayed digits are changing too fast to be
recognized as the power consumption changes.
DISPLAY HELP
If the AMPS or VOLTS displays show the word HELP,
the maximum peak range of the 5300 PA has been
exceeded. This can occur from a high rms input or the
peak of a high crest factor wave form. "HELP" can be
a transient condition resulting from high momentary
starting current and is of no danger if the "HELP"
message disappears after a few seconds. The Auto up-
ranging feature of the 5300 will automatically up range
the instrument as the current or voltage increases over
each ranges maximum, up to the 100 amps range. The
5300 can sustain an overload of up to 200 amps for 5
seconds, maximum. Circuit breaker(s) should be
installed for personal safety and protection of the 5300
PA. See "Current Overload." in Section 1 - Introduction.
POLYPHASE LINE BALANCE
Good polyphase line operating efficiency dictates that
line voltages and currents be closely balanced. Usually,
this requires putting some effort into balancing the loads.
Also, power factor (PF) measurements will be inaccurate
unless a good line balance is maintained.