ABBREVIATIONS
PA - Model 5300 Power Analyzer.
GPIB - IEEE-488 Instrument Bus Standard.
LED - front panel indicator light.
A, V, W, PF and VA: Amperes, Volts, Watts, Power
Factor and Volt Amperes, respectively
MPU - Microprocessor unit.
LSD, LSB - least significant digit or bit.
CMRR - common mode rejection ratio.
UNPACKING
The Model 5300 is packed in reusable, shock resistant
packing material that will protect the instrument during
normal handling. In the event of shipping damage,
immediately notify the carrier and Magtrol Customer
Service Dept. Save all shipping materials for reuse when
returning the instrument for calibration or servicing.
POWER
The 5300 PA is factory wired for either 120 or 240
Vrms, 50/60 Hz, power at 120 volt amps, max. The
line cord is a detachable NEMA standard 3 wire which
plugs into the rear access power entry and filter module
of the 5300 PA. The line cord must be detached from
the PA during servicing.
NOTE: The standard 5300 is factory
wired for 120 Vrms power.
Wiring for 240 Vrms is available
on special order.
FUSE REPLACEMENT
Remove the line cord and carefully pry up and remove
the fuse holder. Use a 2 Amp slow blow fuse for 120
Vrms or a 1 Amp slow blow fuse for 240 Vrms power.
1 - Introduction1 - Introduction
1 - Introduction1 - Introduction
1 - Introduction
SAFETY
Securely ground the 5300 PA case by connecting a good
earth ground at the ground stud that is located on the
rear panel. Use a number 12 AWG wire, or larger.
CONNECTORS
The following plugs mate with 5300 PA connectors:
Superior Electric Supercon
®
INPUT (plug, red, female) - PS100GR
OUTPUT (plug, red, male) - PP100GR
VOLTS HIGH (plug, black, female) - PS25GB
VOLTS LOW (plug, white, female) - PS25GWT
For your convenience, the above connectors are supplied
with your 5300 PA.
Use only the above connectors.
NOTE : GPIB (IEEE-488) 24 Pin bus
standard cable is not included
with the 5300 PA and must be
ordered separately.
MAXIMUM RATINGS
LOAD current: 100 Amperes rms, continuous.
200 Amperes rms, 5 sec. max.
REMOTE VOLTAGE SENSE: Line to line is 600
Vrms. Also, do not exceed 600 Vrms differential
between the VOLTS INPUT and AMPS OUTPUT
terminals.