Magtrol Model 5300 Three Phase Power AnalyzerChapter 3 - Connecting the 5300
SINGLE PHASE AC OR DC
CONNECTIONS
See figure 2.
NOTE: Connections are shown grouped
in cables.
This diagram shows the phase 1 (Ø
1
) voltage and current
inputs connected for measurement. However, any of
the 3-phase inputs can be used. The unused two inputs
are jumpered to insure zero inputs.
Power in watts is calculated as follows:
P = E × I × cos q
( where q is the phase angle between E and I)
On the 5300, select DISPLAY - Ø
1
and
MODE -PH. V.
The 5300 displays VOLTS, AMPS and WATTS.
This circuit uses the 5300 PA remote voltage sense
feature by measuring the voltage at the load. This
increases measurement accuracy by eliminating line
voltage drop from the power measurement. For safety,
an overload circuit breaker (CB) removes all load
voltage during an over-current condition. The voltage
sense lines are connected at the line side of CB to help
prevent inductive transients from entering the PA as
the CB opens. Make sure that connections from CB to
the load are heavy conductors and short as possible.
If a circuit breaker is used in the input line
to the 5300 PA, a circuit should be used
that prevents the breaker from opening until
after the load side breaker has opened.
Otherwise, potentially damaging inductive
transients can be applied to the 5300 PA.
Damage caused by these transients are
outside the scope of the MAGTROL
WARRANTY.
For voltage sense lines less than 25 feet in
length, a twisted pair of #20 gauge (or
larger) wire can be used. For lines longer
than 25 feet or lines grouped with other AC
conductors, shielded cable should be used.
Connect the shield at the 5300 PA GND
terminal. If the wires connecting the load to
the 5300 PA OUTPUT are short and the
resulting voltage drop is insignificant, the
voltage sense connections can be made at
the rear panel of the 5300 PA.
Connect the chassis ground terminal to
a good earth ground. Use at least #12
gauge insulated copper wire.
Figure 2. Single Phase AC or DC Connections
INPUT AMPS OUTPUT VOLTS
Ø1
Ø2
Ø3
HIGHLOW
LOW
(WHITE)
CB
L
O
A
D
EARTH
(BLACK)
HIGH
VOLTAGE SENSE
LOAD
V
LINE =
V
LOAD
1
1
SURGE PROTECTION NOT SHOWN
5300 PA
2
INDICATES CABLING WITH 2 CONDUCTORS.
CABLE MUST BE FULL LOAD
2
2
I
LINE =
I
LOAD
R
R
W B
B
W
W B
R
Ø1
Ø2
Ø3
R
R
R